Sensoface
Display
Problem
Status
Asymmetry
potential and
slope
Asymmetry potential (zero)
and slope of the sensor are
still okay.
The sensor should be replaced
soon.
Asymmetry potential and
slope of the sensor have
reached values which no lon-
ger ensure proper calibration.
Replace sensor.
Calibration
timer
Over 80% of the calibration
interval has already past.
The calibration interval has
been exceeded.
Sensor defect
Check the sensor and its
connections (see also Error
Messages Err 15 and Err 16).
Response
time
Sensor response time has
increased.
The sensor should be replaced
soon.
To achieve an improvement, clean
the sensor and soak it in buffer.
Sensor response time signifi -
cantly increased ( > 72 s, cali-
bration aborted after 120 s)
Replace sensor.