45
Confi guration
Menu item
Action
Choices
Temp detection
during calibration
Select mode using
:
AUTO: Measured by
sensor
MAN: Direct input of
temperature, no measure-
ment (see next step)
EXT: Temperature
specified via current input
(only if TAN E enabled)
Confirm with
enter
AUTO
MAN
EXT
(Manual temperature)
Modify digit using
,
select next digit using
keys.
Confirm with
enter
–20...200 °C
(–4...+392 °F)
Calibration Mode
Select CALMODE using
keys:
AUTO: Calibration with
Calimatic buffer set
recognition
MAN: Manual input of
buffer solutions.
DAT: Input of adjustment
data of premeasured
sensors
Confirm with
enter
AUTO
MAN
DAT
(AUTO: Buff er set)
Select buffer set using
keys (see buf-
fer tables for nominal
values).
Confirm with
enter
-00-...-09-, -U1-
Pressing the
info
key dis-
plays the manufacturer
and nominal values in the
lower line.
5