43
Confi guration
Menu item
Action
Choices
Select sensor type
Select sensor type using
keys.
Confirm with
enter
STANDARD
ISFET
INDUCON (DS)
ISM-DISO (DS)
MEMOSENS (DS)
Temperature unit
Select °C or °F using
keys.
Confirm with
enter
°C
/ °F
Temp detection
during measurement
Select mode using
:
AUTO: Measured by
sensor
MAN: Direct input of
temperature, no measure-
ment (see next step)
EXT: Temperature speci-
fied via current input
(only if TAN E enabled)
Confirm with
enter
AUTO
MAN
EXT
(Manual temperature)
Modify digit using
,
select next digit using
keys.
Confirm with
enter
–20...200 °C
(–4...+392 °F)