Temp Probe Adjustment
Display
Action
Remark
Select temp adjust-
ment. Press
enter
to
proceed.
Wrong settings
change the mea-
surement proper-
ties!
Measure the tempera-
ture of the process me-
dium using an external
thermometer.
Display (3 sec)
Now the device is in
HOLD mode.
Enter the measured
temperature value.
Maximum diff erence:
10 K.
Press
enter
to proceed.
Display of actual
temperature (un-
compensated) in the
lower display.
The corrected tempera-
ture value is displayed.
Sensoface is active.
To end calibration:
Select MEAS, then
enter
To repeat calibration:
Select REPEAT, then
enter
After calibration is
ended, the device will
switch to measuring
mode.
After end of calibra-
tion, the outputs re-
main in HOLD mode
for a short time.
Temp probe adjustment