31
Alarm
Manual Activation of HOLD
The HOLD can be activated manually from the HOLD menu.
This allows checking or replacing a sensor, for example, without
provoking unintended reactions of outputs or contacts.
Press
meas
key to return to selection menu.
Alarm
When an error has occurred,
Err xx
is displayed immediately.
Only after expiry of a user-defi ned delay time will the alarm be
registered and entered in the logbook.
During an alarm the display blinks, the display backlighting turns
red
.
Error messages can also be signaled by a 22 mA output current
(see Confi guration).
The alarm contact is activated by alarm and power failure, see also
“Confi guration / Alarm Settings“.
2 sec after the failure event is corrected, the alarm status will be
deleted.
External Activation of HOLD
The HOLD mode can be activated from outside by sending a signal to
the Hold input (e.g. from the process control system).
10
11
Process control system
Power supply
12...24 V AC/DC
Stratos Pro A4...
Input
HOLD
HOLD inactive
0...2 V AC/DC
HOLD active
10...30 V AC/DC