![Knick Stratos Pro A4..MSPH series Instruction Manual Download Page 89](http://html.mh-extra.com/html/knick/stratos-pro-a4-msph-series/stratos-pro-a4-msph-series_instruction-manual_1980197089.webp)
89
Display
Action
Remark
At the end of the stabil-
ity check, the value will
be saved and the asym-
metry potential will be
displayed.
Calibration with the fi rst
buff er is terminated.
Remove the sensor and
temp probe from the
fi rst buff er solution and
rinse them thoroughly.
Use the arrow keys to
select:
END (1-point cal)
•
CAL2 (2-point cal)
•
REPEAT
•
Press
enter
to proceed.
Please note:
Stability check can
be stopped after
10 sec (by pressing
enter
). However, this
reduces calibration
accuracy. Display for
1-point cal:
Sensoface is active.
End with
enter
2-point calibration:
Immerse sensor and
temperature probe in
the second buff er solu-
tion. Start with
enter
The calibration pro-
cess runs as for the
fi rst buff er.
Retract sensor and
temp probe out of
second buff er, rinse off ,
re-install.
Press
enter
to proceed.
The slope and asym-
metry potential of
the sensor (based on
25 °C) are displayed.
Use the arrow keys to
select:
MEAS (end)
•
REPEAT
•
Press
enter
to proceed.
End: HOLD is deacti-
vated with delay.
When 2-point cal is
ended:
Automatic Calibration (Calimatic)
Summary of Contents for Stratos Pro A4..MSPH series
Page 1: ...www knick de Latest Product Information Stratos Pro A4 MSPH Instruction Manual ...
Page 20: ...20 ...
Page 68: ...68 ...
Page 111: ...111 ...
Page 147: ......