Error Handling
Alarm condition:
The display backlighting turns
•
red
The alarm icon
•
is displayed
The complete measured-value display blinks
•
„
•
ERR xxx
“ is displayed in the lower menu line
Press the [
info
] key to view a short error text:
The error text appears in the lower menu line
•
The main display reads “
•
InFo
“.
Parameter errors:
Confi guration data such as current range, limit values, etc are checked
during the input.
If they are out of range,
“
•
ERR xxx
“ is displayed for 3 sec,
the display backlighting fl ashes red,
•
the respective maximum or minimum value is shown,
•
input must be repeated.
•
If a faulty parameter arrives through the interface (IrDA, HART),
an error message will be displayed: “
•
ERR 100...199
“
the faulty parameter can be localized by pressing the [
•
info
] key
Calibration errors:
If errors occur during calibration, e.g. by using a wrong calibration
solution,
an error message will be displayed
•
calibration will be restarted
•
Sensoface:
If the Sensoface becomes sad,
the display backlighting will turn purple
•
the cause can be seen by pressing the
•
info
key
the calibration data can be seen in the Diagnostics menu
•