63
Confi guration
Menu item
Action
Choices
Use of relays
Select in the text line
using
keys:
Limit function (LIMITS)
•
Controller
•
(CONTROLLER)
Confirm with
enter
LIMITS
/ CONTROLLER
Please note:
Selecting
CONTROLLER leads to
Controller menu group
CTR.
Select process
variable
Select desired process
variable using
keys.
Confirm with
enter
PH
/ORP/TMP
Limit 1 function
Select desired function
using arrow keys.
LoLevel: active if value
falls below / HiLevel:
active if value exceeds
setpoint
Confirm with
enter
Lo LEVL
/ Hi LEVL
Limit 1
contact response
N/O: normally open
contact
N/C: normally closed
contact
Select using
keys.
Confirm with
enter
N/O
/ N/C
Limit 1 setpoint
Enter setpoint using
keys.
Confirm with
enter
–2.00...16.00 PH
(
00.00 PH
) /
-1999...1999 mV /
–20...200 °C
5