37
Confi guration
Confi guration
Choices
Default
Temperature compensation (CORRECTION)
COR:
TC LIQUID
–19.99...19.99%/K
00.00%/K
TEMP EXT
*
ON/OFF
OFF
ON
I-INPUT
0...20 mA/4...20 mA
4...20 mA
°C
BEGIN 4 mA –20...200 °C
000.0 °C
END 20 mA –20...200 °C
100.0 °C
°F
BEGIN 4 mA –4...392 °F
032.0 °F
END 20 mA –4...392 °F
212.0 °F
Alarm (ALARM)
ALA:
DELAYTIME
0...600 SEC
0010 SEC
SENSOCHECK
ON/OFF
OFF
Relay 1/2 (RL1/RL2)
REL:
(Selected in text line)
LIMITS, CONTROLLER
LIMITS
RL1:
CHANNEL
PH/ORP/TMP
PH
FUNCTION
Lo LEVL, Hi LEVL
Lo LEVL
CONTACT
N/O, N/C
N/O
LEVEL
–2.00...16 PH
(-1999...1999 mV)
(-20...200 °C)
00.00 PH
HYSTERESIS
0...10.00 PH
00.50 PH
DELAYTIME
0...9999 SEC
0010 SEC
RL2
CHANNEL
PH/ORP/TMP
PH
FUNCTION
Lo LEVL, Hi LEVL
Hi LEVL
CONTACT
N/O, N/C
N/O
LEVEL
–2.00...16 PH
(-1999...1999 mV)
(-20...200 °C)
14.00 PH
HYSTERESIS
0...10.00 PH
00.50 PH
DELAYTIME
0...9999 SEC
0010 SEC
*) is only displayed if enabled and SENSOR TEMP EXT has been selected.