Choices
CIP / SIP
The following adjustments are possible for INDUCON digital sensors:
Cleaning cycle
counter
Select ON or OFF using
keys.
Activates/deactivates log-
ging in extended logbook
and counters (if provided)
Confirm with
enter
ON/
OFF
Cleaning cycles
Only with CIP COUNT ON:
Enter value using
keys.
Confirm with
enter
0...9999 CYC
(
0025 CYC
)
Not with Memosens
Sterilization cycle
counter
Select ON or OFF using
keys.
Activates/deactivates log-
ging in extended logbook
and counters (if provided)
Confirm with
enter
ON/
OFF
Sterilization cycles
Only with SIP COUNT ON:
Enter value using
keys.
Confirm with
enter
0...9999 CYC
(
0025 CYC
)
Not with Memosens
The cleaning and sterilization cycles are counted to measure the load
on the sensor.
Suitable for biochemical applications (process temp approx. 0...50 °C,
CIP temperature > 55 °C, SIP temperature > 115 °C).