Precautions for high-speed load simulations
During high-speed load simulation, do not connect the common terminal of the external
device to the terminal of the DUT (the terminal that is connected to the negative input terminal
of the PLZ-4WL). Attach a commercially available ferrite core to the wiring between the PLZ-
4WL and the external device.
22
STATUS COM
STATUS signal common for pins 17 to 20.
23
N.C.
---
24
N.C.
---
25
SHORT SIGNAL OUT
Relay contact output (30 VDC/1 A)
26
SHORT SIGNAL OUT
1
Valid only when the front panel setting is H range.
2
The maximum applied voltage of the photocoupler is 30 V, and the maximum current is 8 mA.
3
Pin
No.
Signal Name
Description
RANGE CONT 0
RANGE CONT 1
H range
H
H
M range
H
L
L range
L
H
RANGE STATUS 0
RANGE STATUS 1
H range
OFF
OFF
M range
OFF
ON
L range
ON
OFF
DUT
L
L
PLZ-4WL Series
ACOM
External control terminal
External
device
Load input
terminals
+
-
Do not connect
Attach a ferrite core