PLZ-4WL
75
Sequence Example (Fast Sequence)
4
Sequenc
es
Editing program 11
Memo
You can move the cursor
by pressing
,
,
,
and .
To exit the sequence
editing screens, press
PREV (SHIFT+
).
1
Press EDIT (SHIFT+A).
The program editing screen appears.
2
Set the program number to 11.
If the cursor is not blinking at No., press
and
to move it
there. Turn the rotary knob to select the number.
3
Press to move the cursor to Memo, and enter
“Program11.”
Enter the text using the rotary knob and and .
4
Press to move the cursor to Mode, and select an
operation mode (CC mode).
Turn the rotary knob to select FCC.
5
Press to move the cursor to Range, and set the
current and voltage ranges to 50 A and 4 V.
Turn the rotary knob to select 50 A, 4 V.
6
Press to move the cursor to Loop, and set the
number of times to execute the program to 3.
Turn the rotary knob to select the number.
7
Press to move the cursor to Last Load, and
select the state that the load will be in after the
sequence ends (select ON).
Turn the rotary knob to select ON.
8
Press to move the cursor to Last Set, and select
the current that the load will supply after the
sequence ends (select 1 A).
Turn the rotary knob to select the number.
9
Press to move the cursor to RPTSTEP, and set
the step number of the last step to 15.
Turn the rotary knob to select the number.
10
Press to move the cursor to TIME BASE, and set
the step execution time to 1 ms.
Turn the rotary knob to set the time.
11
Press NEXT (SHIFT+ ) or ENTER.
The step value screen appears.
SEQ
PREV
NEXT
No. :
11
Memo :
Mode : FCC
Loop : 0003
Ran
g
e : 50A , 4V
SEQ
PREV
NEXT
Last Load : ON
Last Set : 1.000
A
RPTSTEP : 15
TIME BASE : 1ms