PLZ-4WL
55
Menu
3
Advance
d Oper
ati
ons
List of menu items
(the factory default settings are underlined)
Item 1
Item 2
Item 3
Settings
Description
1.Setup
1.Function
Soft Start
OFF:Disable soft start
100µs/200µs/500µs/1ms/2ms/5ms/
10ms/20ms
Soft start time
Count Time
OFF:Not displayed
ON:Displayed
Elapsed time display
2.Protect Action
OCP
LOAD OFF: The load is turned off.
LIMIT: The current is limited.
Action to be performed when
OCP is activated
OPP
LOAD OFF: The load is turned off.
LIMIT: The current is limited.
Action to be performed when
OPP is activated
3.Memory
Recall
DIRECT
SAFETY
Method for recalling the preset
memory
4.Cut Off
Time
OFF
0:00:01 to 999:59:59
(Hour:minute:second)
Auto load-off timer
Press and to move the
cursor between the hour, minute,
and second positions.
5.Response
CV Response
FAST
NORMAL
CV mode response speed
CR Response
FAST
NORMAL
CR mode response speed
2.Configuration
1
1
The Configuration conditions, except for remote sensing, are enabled after the power switch is turned off and then on again.
1.Power On
Load On
OFF: The load is off when the power
is turned on.
ON: The load is on when the power is
turned on.
State of the load when the power
switch is turned on
Key Lock
OFF: Keys are not locked.
ON: Keys are locked.
Key lock
2.Interface
Control
GPIB/ RS232C/ USB
Interface setting
GPIB
Address 1 to 30 (1)
GPIB address
RS232C Baudrate 2400bps/4800bps/9600bps/19200bps Baud rate
Data
2
,
Stop
2
Information about the PLZ-4WL. It cannot be changed.
8,
1/2
Data length (fixed at 8 bits) and
the stop bit
Parity
2
NONE
Parity (fixed at NONE)
Ack
OFF, ON
Acknowledgment
USB
2
VID
0x0B3E
Vendor ID
PID
0x1019(PLZ164WL)
0x1020(PLZ334WL)
Product ID
S/N
AB123456 (example)
Serial number
3.External
CC/CR/CP
OFF: Disabled, ON: Enabled
External control for the CC, CR,
and CP modes
CV
OFF: Disabled, ON: Enabled
External control for CV mode
ADD CC
OFF: Disabled, ON: Enabled
Superimpose current on the CC
mode panel setting through
external control
LoadOn IN
LOW: Turned on by a low level
HIGH: Turned on by a high level
External control logic setting for
turning the load on and off
4.Remote Sensing Sensing
OFF, ON
Remote sensing function
3.Calibration
1.CC(Low)
Calibration of the PLZ-4WL.
For details, see "Calibration" on page 98.
2.CC(Mid)
3.CC(High)
4.CV 4V
5.CV 30V
4.Model Info
2
(MODEL)
PLZxxxWL
Model name
VERSION SUB
x.xx
Firmware version
VERSION MAIN
x.xx
ROM version