104
PLZ-4WL
Ending Calibration
Calibrating the high range
Calibrate the High range in the same way that you calibrated the Low range. In step 4 , select
“5. CV 30 V.”
The calibration of CV mode is complete after you calibrate the High range
.
Ending Calibration
1
Press PREV (SHIFT+ ).
The screen that was displayed before the calibration screen appears.
2
Press MENU (SHIFT+SET/VSET).
The screen that was displayed before the menu appears.
When you press ENTER or NEXT (SHIFT+ ) in the gain calibration screen, the calibration
data is written to the internal memory.
If you only want to check the calibration data, be sure to press PREV (SHIFT+ ) or MENU
(SHIFT+SET/VSET) to exit the gain calibration screen.