112
PLZ-4WL
Specifications
Soft start
Response
Remote sensing
Protection function
Model
PLZ164WL
PLZ334WL
Operation mode
CC
Selectable times
1
1
The time it takes to shift from 10 % to 90 % of the rated current
OFF, 100 µs, 200 µs, 500 µs,
1 ms, 2 ms, 5 ms, 10 ms, or 20 ms
Time accuracy
±(30 % of set +10 µs)
Model
PLZ164WL
PLZ334WL
Response speed
1
1
Can be set in CV mode and CR mode
NORMAL, FAST
Model
PLZ164WL
PLZ334WL
Voltage that can be compensated
3 V for a single line
Model
PLZ164WL
PLZ334WL
Overvoltage protection (OVP) Turns off the load at 115 % of the rated voltage
Overcurrent protection (OCP)
Can be set to a value from 10 % to 110 % of the rated current
The action can be set to load off or limit.
Overpower protection (OPP)
Can be set to a value from 10 % to 110 % of the rated power
The action can be set to load off or limit.
Overheat protection (OHP)
Turns off the load when the heat sink temperature reaches 90
C
Undervoltage protection
(UVP)
Turns off the load when the specified value is detected
Can be set to a value from 0.3 V to 30 V
Reverse-connection
protection (REV)
Implemented through a fuse
Turns off the load when an alarm occurs