PLZ-4WL
65
Sequence Example (Normal Sequence)
4
Sequenc
es
Editing program 1
In this example, program 1 will be created first and then program 2. You can edit the
programs in any order that you like. It is OK to create program 2 first and then create program
1.
Before you set the steps for program 1, you can also configure program 2.
Memo
You can move the cursor
by pressing
,
,
,
and .
To exit the sequence
editing screens, press
PREV (SHIFT+
).
1
Press EDIT (SHIFT+A).
The program editing screen appears.
2
Set the program number to 1.
If the cursor is not blinking at No., press
and
to move it
there. Turn the rotary knob to select the number.
3
Press to move the cursor to Memo, and enter
“Program1.”
Enter the text using the rotary knob and and
.
4
Press to move the cursor to Mode, and select an
operation mode (CC mode).
Turn the rotary knob to select NCC.
5
Press to move the cursor to Range, and set the
current and voltage ranges (50 A and 30 V).
Turn the rotary knob to select 50 A, 30 V.
6
Press to move the cursor to Loop, and set the
number of times to execute the program to 1.
Turn the rotary knob to select the number.
7
Press to move the cursor to Last Load, and
select the state that the load will be in after the
sequence ends (select OFF).
In this example, program 2 will be executed after program 1,
so this setting is ignored.
8
Press to move the cursor to Last Set, and select
the current that the load will supply after the
sequence ends (select 0 A).
In this example, program 2 will be executed after program 1,
so this setting is ignored.
9
Press to move the cursor to Chain, and set the
number of the program to execute next to 2.
To execute program 2 next, select 2.
10
Press NEXT (SHIFT+ ) or ENTER.
The step value screen appears.
SEQ
PREV
NEXT
No. :
1
Memo :
Mode :
NCC
Loop :
0001
Ran
g
e : 50A, 30V
SEQ
PREV
NEXT
No. :
1
Memo :
Mode :
NCC
Loop :
0001
Ran
g
e : 50A, 30V
Program1
SEQ
PREV
NEXT
Last Load
: OFF
Last Set
: 0.000
A
Chain :