PLZ-4WL
63
Editing Steps in a Normal Sequence
4
Sequenc
es
Step-execution pattern
Set the execution pattern.
■
RAMP
Use this setting to specify the step transition. If you set RAMP to ON, the current changes in
a slope within the step; if you set it to OFF, the current changes instantly and is sustained
throughout the step.
■
TRIG
Use this setting to specify whether or not to produce a trigger signal. If you set TRIG to ON,
the TRIG OUT terminal on the front panel produces a trigger signal when the step is
executed.
■
PAUSE
Use this setting to specify whether or not to pause the program. If you set PAUSE to ON, the
sequence is paused after the step is executed. To unpause the sequence, press PAUSE
(SHIFT+C) on the front panel, or apply a trigger signal to the TRIG INPUT pin (pin 13) of the
J1 connector.
Deleting steps
Press
and
to highlight the step that you want to delete, and then press DEL
(SHIFT+
). The highlighted step is deleted.
Item
Setting
Description
LOAD
ON: Load on, OFF: Load off
Turns the load on or off.
RAMP
ON: Slope, OFF: Step
The ramp (step transition)
TRIG
ON: Output, OFF: Don’t output
The trigger signal during step execution
PAUSE
ON: Pause, OFF: Don’t pause
Whether or not to pause
Example: Setting: 10 A
Step transition (RAMP OFF)
Step execution time: 1 s
Example: Setting: 10 A
Ramp transition (RAMP ON)
Step execution time: 1 s
1 s
10 A
1 s
10 A
Step n - 1
Step n
TRIG OUT
Example: Set trigger output at step number n
(TRIG ON).
See
Trigger signal input
(TRIG IN)
PAUSE key
Program in
execution
Paused
Paused
Program execution
resumed
Program execution
resumed
at step n
PAUSE ON
PAUSE
PAUSE key
PAUSE
Example: Pause and resume using the
PAUSE (SHIFT+C) key.
Example: Pause by executing
PAUSE in a program, and
resume using trigger input
(TRIG IN).
(SHIFT+C)
(SHIFT+C)