108
PLZ-4WL
Specifications
Unless specified otherwise, the specifications are for the following settings and conditions.
•
The warm-up time is 30 minutes (with current flowing).
•
TYP:
These are typical values that are representative of situations where the
PLZ-4WL operates in an environment with an ambient temperature of 23
C.
These values do not guarantee the performance of the PLZ-4WL.
•
After PLZ-4WL has been warmed up, it must be calibrated correctly in a 23
C ± 5
C
environment according to the procedures given in the operation manual.
•
% of set: Denotes a percentage of the input voltage, input current, or input power
setting.
•
% of f.s: Denotes a percentage of the rated input voltage, rated input current, or rated
input power.
•
% of reading: Denotes a percentage of the input voltage, input current, or input power
reading.
Ratings
Constant current (CC) mode
Model
PLZ164WL
PLZ334WL
Operating voltage (DC)
1
1
The minimum operating voltage at which current begins to flow through the PLZ-4WL is approximately
50 mV. At the load input terminals.
0.3 V to 30 V
2
2
The minimum operating voltage (including the voltage drop due to the wire inductance component) in
switching mode increases by approximately 40 mV per 1 A/µs of the slew rate setting.
Current
50 A
100 A
Power
165 W
330 W
Model
PLZ164WL
PLZ334WL
Operating range
H range
0 A to 50 A
0 A to 100 A
M range
0 A to 5 A
0 A to 10 A
L range
0 A to 500 mA
0 A to 1 A
Setting range
H range
0 A to 52.5 A
0 A to 105 A
M range
0 A to 5.25 A
0 A to 10.5 A
L range
0 A to 525 mA
0 A to 1.05 A
Resolution
H range
2 mA
5 mA
M range
0.2 mA
0.5 mA
L range
0.02 mA
0.05 mA
Accuracy of setting
±(0.2 % of set + 0.1 % of f.s
1
) + Vin
2
/150 k
1
The full scale of the range. However, for the M range, it is the full scale of the H range.
2
Vin: The voltage at the load input or sensing terminals
Input voltage variation
3
3
When the input voltage is changed from 0.3 V to 30 V at a current equal to the rated power/30 V.
±(0.1 % of set + 0.02 % of f.s
Ripple
rms
4
4
Measurement frequency bandwidth: 10 Hz to 1 MHz
4 mA
8 mA
p-p
5
5
Measurement frequency bandwidth: 10 Hz to 20 MHz
40 mA
80 mA