9
SENSe Subsystem
[SENSe:]LRESistance:RESolution
218
U3606A Programmer’s Reference
Example
See also
“CONFigure:LRESistance” on page 104
&
LRES:RES 0.0001
This command sets the measurement
resolution to 100
μΩ
.
&
LRES:RES?
This query returns the resolution selected.
$
+1.000000E-04
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply Programmer s Reference ...
Page 2: ......
Page 3: ......
Page 14: ...XII U3606A Programmer s Reference ...
Page 32: ...1 Introduction to SCPI SCPI Status System 18 U3606A Programmer s Reference ...
Page 58: ...2 CALCulate Subsystem CALCulate NULL OFFSet 44 U3606A Programmer s Reference ...
Page 92: ...3 CALibration Subsystem Remote Calibration Procedures 78 U3606A Programmer s Reference ...
Page 132: ...4 CONFigure Subsystem CONFigure DCYCle 118 U3606A Programmer s Reference ...
Page 138: ...5 INITiate Subsystem INITiate CONTinuous 124 U3606A Programmer s Reference ...
Page 174: ...7 MEMory Subsystem MEMory STATe RECall AUTO 160 U3606A Programmer s Reference ...
Page 295: ...SOURce Subsystem 10 SOURce PROTection STATe U3606A Programmer s Reference 281 RST on page 328 ...
Page 296: ...10 SOURce Subsystem SOURce PROTection STATe 282 U3606A Programmer s Reference ...
Page 332: ...14 Root Commands READ 318 U3606A Programmer s Reference ...
Page 350: ...15 IEEE 488 2 Common Commands WAI 336 U3606A Programmer s Reference ...
Page 370: ...Command Quick Reference 356 U3606A Programmer s Reference ...