CALibration Subsystem
3
Remote Calibration Procedures
U3606A Programmer’s Reference
77
Repeat the voltage output adjustment procedures again for the rear output
terminals (send the
SOUR:SENS EXT
command). See the
U3606A User’s
and Service Guide
for more information on how to connect the load leads
to the rear terminal block.
Voltage output adjustment — S1 (30 V/1 A) range
14
Select the CV output, 30 V range.
&
SOUR:VOLT:RANG 30
15
Calibrate the lower point for CV output.
&
CAL:LEV MIN
16
Calibrate the upper point for CV output.
&
CAL:LEV MAX
17
Connect an additional 30
Ω
, 50 W load
across the
(red) and
(black) output
terminal. Leave the connections from the
output terminals to the input terminals
intact.
18
Calibrate the load point for CV output.
&
CAL:LEV LOAD
Calibration steps
Remote commands
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply Programmer s Reference ...
Page 2: ......
Page 3: ......
Page 14: ...XII U3606A Programmer s Reference ...
Page 32: ...1 Introduction to SCPI SCPI Status System 18 U3606A Programmer s Reference ...
Page 58: ...2 CALCulate Subsystem CALCulate NULL OFFSet 44 U3606A Programmer s Reference ...
Page 92: ...3 CALibration Subsystem Remote Calibration Procedures 78 U3606A Programmer s Reference ...
Page 132: ...4 CONFigure Subsystem CONFigure DCYCle 118 U3606A Programmer s Reference ...
Page 138: ...5 INITiate Subsystem INITiate CONTinuous 124 U3606A Programmer s Reference ...
Page 174: ...7 MEMory Subsystem MEMory STATe RECall AUTO 160 U3606A Programmer s Reference ...
Page 295: ...SOURce Subsystem 10 SOURce PROTection STATe U3606A Programmer s Reference 281 RST on page 328 ...
Page 296: ...10 SOURce Subsystem SOURce PROTection STATe 282 U3606A Programmer s Reference ...
Page 332: ...14 Root Commands READ 318 U3606A Programmer s Reference ...
Page 350: ...15 IEEE 488 2 Common Commands WAI 336 U3606A Programmer s Reference ...
Page 370: ...Command Quick Reference 356 U3606A Programmer s Reference ...