CALibration Subsystem
3
Remote Calibration Procedures
U3606A Programmer’s Reference
71
AC current gain adjustment — 1 A
10
Select the 1 A range. Input 1 A, 1 kHz AC
current to the
I
(red) and
LO
(black) input
terminals.
&
CONF:CURR:AC 1
11
Calibrate the 1 A point for AC current
measurements.
&
CAL:VAL 1
12
Start the calibration.
&
CAL?
Calibration steps
Remote commands