CALibration Subsystem
3
Remote Calibration Procedures
U3606A Programmer’s Reference
75
Output adjustments
The instrument calculates and stores output corrections for each output
level. The U3606A implements a closed loop output calibration procedure
to its inherent dual function ability as a digital multimeter and a DC
power supply. The output constant is computed from the calibration level
set for the calibration command and from measurements made
automatically during the adjustment procedure.
Adjustments for each function should be performed ONLY in the order
shown.
Calibration steps
Remote commands
1
Place the instrument in the free-run mode.
Measurements are made continuously and
the readings are stored in the instrument
memory.
&
INIT:CONT ON
2
Set the signal source to read the sense
signal inputs from the front output terminals
of the instrument. The U3606A needs to be
calibrated twice. Once for the internal sense
source and again for the external sense
source.
&
SOUR:SENS INT
3
Perform the current and voltage output
adjustments for the front output terminals.
See
4
Set the signal source to read the sense
signal inputs from the rear output terminals
of the instrument. When you have
completed the voltage and current output
adjustments for the internal sense source,
repeat the entire adjustment procedure
again for the external sense source.
&
SOUR:SENS EXT
5
Perform current and voltage output
adjustments for the rear output terminals.
See
6
Remove all connections from the
instrument. Reset the calibration message
and record the new calibration count.
&
CAL:STR "<new_message>"
&
CAL:COUN?
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply Programmer s Reference ...
Page 2: ......
Page 3: ......
Page 14: ...XII U3606A Programmer s Reference ...
Page 32: ...1 Introduction to SCPI SCPI Status System 18 U3606A Programmer s Reference ...
Page 58: ...2 CALCulate Subsystem CALCulate NULL OFFSet 44 U3606A Programmer s Reference ...
Page 92: ...3 CALibration Subsystem Remote Calibration Procedures 78 U3606A Programmer s Reference ...
Page 132: ...4 CONFigure Subsystem CONFigure DCYCle 118 U3606A Programmer s Reference ...
Page 138: ...5 INITiate Subsystem INITiate CONTinuous 124 U3606A Programmer s Reference ...
Page 174: ...7 MEMory Subsystem MEMory STATe RECall AUTO 160 U3606A Programmer s Reference ...
Page 295: ...SOURce Subsystem 10 SOURce PROTection STATe U3606A Programmer s Reference 281 RST on page 328 ...
Page 296: ...10 SOURce Subsystem SOURce PROTection STATe 282 U3606A Programmer s Reference ...
Page 332: ...14 Root Commands READ 318 U3606A Programmer s Reference ...
Page 350: ...15 IEEE 488 2 Common Commands WAI 336 U3606A Programmer s Reference ...
Page 370: ...Command Quick Reference 356 U3606A Programmer s Reference ...