IEEE-488.2 Common Commands
15
*TRG
U3606A Programmer’s Reference
333
*TRG
Syntax
*TRG
The
*TRG
(TRiGger) is used in conjunction with the
TRIGger:SOURCe
command to trigger the instrument from the remote interface.
Remarks
•
Use the
TRIGger:SOURce
command to select the
BUS
(software) trigger
source.
•
After setting the trigger source, you must place the multimeter in the
“wait- for- trigger” state using the
INITiate[:IMMediate]
command.
The
*TRG
command will not be accepted unless the multimeter is in the
“wait- for- trigger” state.
Example
See also
“INITiate[:IMMediate]” on page 120
&
TRIG:SOUR BUS
This command selects the bus (software)
trigger source.
&
INIT
This command places the instrument in the
"wait-for-trigger" state.
&
*TRG
This command triggers the instrument.
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply Programmer s Reference ...
Page 2: ......
Page 3: ......
Page 14: ...XII U3606A Programmer s Reference ...
Page 32: ...1 Introduction to SCPI SCPI Status System 18 U3606A Programmer s Reference ...
Page 58: ...2 CALCulate Subsystem CALCulate NULL OFFSet 44 U3606A Programmer s Reference ...
Page 92: ...3 CALibration Subsystem Remote Calibration Procedures 78 U3606A Programmer s Reference ...
Page 132: ...4 CONFigure Subsystem CONFigure DCYCle 118 U3606A Programmer s Reference ...
Page 138: ...5 INITiate Subsystem INITiate CONTinuous 124 U3606A Programmer s Reference ...
Page 174: ...7 MEMory Subsystem MEMory STATe RECall AUTO 160 U3606A Programmer s Reference ...
Page 295: ...SOURce Subsystem 10 SOURce PROTection STATe U3606A Programmer s Reference 281 RST on page 328 ...
Page 296: ...10 SOURce Subsystem SOURce PROTection STATe 282 U3606A Programmer s Reference ...
Page 332: ...14 Root Commands READ 318 U3606A Programmer s Reference ...
Page 350: ...15 IEEE 488 2 Common Commands WAI 336 U3606A Programmer s Reference ...
Page 370: ...Command Quick Reference 356 U3606A Programmer s Reference ...