CALibration Subsystem
3
Remote Calibration Procedures
U3606A Programmer’s Reference
73
Capacitance gain adjustment — 10 nF
10
Select the 10 nF range. Input 10 nF to the
(red) and
LO
(black) input terminals.
&
CONF:CAP 10n
11
Calibrate the 10 nF point for capacitance
measurements.
&
CONF:VAL 1E-8
12
Start the calibration.
&
CAL?
Capacitance gain adjustment — 100 nF
13
Select the 100 nF range. Input 100 nF to the
(red) and
LO
(black) input terminals.
&
CONF:CAP 100n
14
Calibrate the 100 nF point for capacitance
measurements.
&
CONF:VAL 1E-7
15
Start the calibration.
&
CAL?
Capacitance gain adjustment — 1
μ
F
16
Select the 1
μ
F range. Input 1
μ
F to the
(red) and
LO
(black) input terminals.
&
CONF:CAP 1u
17
Calibrate the 1
μ
F point for capacitance
measurements.
&
CONF:VAL 1E-6
18
Start the calibration.
&
CAL?
Capacitance gain adjustment — 10
μ
F
19
Select the 10
μ
F range. Input 10
μ
F to the
(red) and
LO
(black) input terminals.
&
CONF:CAP 10u
20
Calibrate the 10
μ
F point for capacitance
measurements.
&
CONF:VAL 1E-5
21
Start the calibration.
&
CAL?
Capacitance gain adjustment — 100
μ
F
22
Select the 100
μ
F range. Input 100
μ
F to the
(red) and
LO
(black) input terminals.
&
CONF:CAP 100u
23
Calibrate the 100
μ
F point for capacitance
measurements.
&
CONF:VAL 1E-4
24
Start the calibration.
&
CAL?
Calibration steps
Remote commands
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply Programmer s Reference ...
Page 2: ......
Page 3: ......
Page 14: ...XII U3606A Programmer s Reference ...
Page 32: ...1 Introduction to SCPI SCPI Status System 18 U3606A Programmer s Reference ...
Page 58: ...2 CALCulate Subsystem CALCulate NULL OFFSet 44 U3606A Programmer s Reference ...
Page 92: ...3 CALibration Subsystem Remote Calibration Procedures 78 U3606A Programmer s Reference ...
Page 132: ...4 CONFigure Subsystem CONFigure DCYCle 118 U3606A Programmer s Reference ...
Page 138: ...5 INITiate Subsystem INITiate CONTinuous 124 U3606A Programmer s Reference ...
Page 174: ...7 MEMory Subsystem MEMory STATe RECall AUTO 160 U3606A Programmer s Reference ...
Page 295: ...SOURce Subsystem 10 SOURce PROTection STATe U3606A Programmer s Reference 281 RST on page 328 ...
Page 296: ...10 SOURce Subsystem SOURce PROTection STATe 282 U3606A Programmer s Reference ...
Page 332: ...14 Root Commands READ 318 U3606A Programmer s Reference ...
Page 350: ...15 IEEE 488 2 Common Commands WAI 336 U3606A Programmer s Reference ...
Page 370: ...Command Quick Reference 356 U3606A Programmer s Reference ...