SENSe Subsystem
9
[SENSe:]LRESistance:RANGe:AUTO
U3606A Programmer’s Reference
215
[SENSe:]LRESistance:RANGe:AUTO
Syntax
[SENSe:]LRESistance:RANGe:AUTO <mode>
This command disables or enables autoranging for 4- wire low- resistance
measurements. Autoranging is convenient because the instrument
automatically selects the range for each measurement based on the input
signal detected.
[SENSe:]LRESistance:RANGe:AUTO?
This query returns a boolean value that represents the 4- wire
low- resistance autoranging setting:
0
or
1
Parameter
Remarks
•
Autorange thresholds:
Down range at: <10% of range
Up range at: >120% of range
•
With autoranging enabled, the instrument selects the appropriate range
based on the input signal detected.
•
Selecting a discrete range (see
“[SENSe:]LRESistance:RANGe[:UPPer]” on page 213
) will disable
autoranging.
•
The instrument enables autoranging after a Factory Reset (
*RST
command).
•
For measuring low- resistance a delayed response should be expected
from the front panel display. For remote interface operation, increase
the SCPI query timeout value. (Typically 15000 ms.)
Item
Type
Range of values
Default value
auto
Boolean
0|1|OFF|ON
1
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply Programmer s Reference ...
Page 2: ......
Page 3: ......
Page 14: ...XII U3606A Programmer s Reference ...
Page 32: ...1 Introduction to SCPI SCPI Status System 18 U3606A Programmer s Reference ...
Page 58: ...2 CALCulate Subsystem CALCulate NULL OFFSet 44 U3606A Programmer s Reference ...
Page 92: ...3 CALibration Subsystem Remote Calibration Procedures 78 U3606A Programmer s Reference ...
Page 132: ...4 CONFigure Subsystem CONFigure DCYCle 118 U3606A Programmer s Reference ...
Page 138: ...5 INITiate Subsystem INITiate CONTinuous 124 U3606A Programmer s Reference ...
Page 174: ...7 MEMory Subsystem MEMory STATe RECall AUTO 160 U3606A Programmer s Reference ...
Page 295: ...SOURce Subsystem 10 SOURce PROTection STATe U3606A Programmer s Reference 281 RST on page 328 ...
Page 296: ...10 SOURce Subsystem SOURce PROTection STATe 282 U3606A Programmer s Reference ...
Page 332: ...14 Root Commands READ 318 U3606A Programmer s Reference ...
Page 350: ...15 IEEE 488 2 Common Commands WAI 336 U3606A Programmer s Reference ...
Page 370: ...Command Quick Reference 356 U3606A Programmer s Reference ...