SENSe Subsystem
9
[SENSe:]LRESistance:RANGe[:UPPer]
U3606A Programmer’s Reference
213
[SENSe:]LRESistance:RANGe[:UPPer]
Syntax
[SENSe:]LRESistance:RANGe[:UPPer] {<range>|MIN|MAX|DEF}
This command selects the measurement range for 4- wire low- resistance
measurements.
[SENSe:]LRESistance:RANGe[:UPPer]? [{MIN|MAX}]
This query returns a numeric value that represents the 4- wire
low- resistance measurement range.
Parameter
Remarks
•
Selecting a discrete range will disable autoranging.
•
If the input signal is greater than can be measured on the selected
range (manual ranging), the instrument gives an overload indication.
•
Positive overload: “
+9.37
” or “OL” from the front panel.
•
Negative overload: “
-9.37
” or “–OL” from the front panel.
•
If a range change is in progress (due to the autoranging setting) when
you query the instrument, the value “
+9.37
” will be
returned. Wait for the instrument to select an appropriate range before
querying the instrument again.
•
The instrument is set to 100 m
Ω
, with autoranging enabled, after a
Factory Reset (
*RST
command).
•
For measuring low- resistance a delayed response should be expected
from the front panel display. For remote interface operation, increase
the SCPI query timeout value. (Typically 15000 ms.)
Item
Type
Range of values
Default value
range
Numeric
•
100 m
Ω
|MIN
•
1000 m
Ω
•
10
Ω
|MAX
•
DEF (AUTO)
MIN
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply Programmer s Reference ...
Page 2: ......
Page 3: ......
Page 14: ...XII U3606A Programmer s Reference ...
Page 32: ...1 Introduction to SCPI SCPI Status System 18 U3606A Programmer s Reference ...
Page 58: ...2 CALCulate Subsystem CALCulate NULL OFFSet 44 U3606A Programmer s Reference ...
Page 92: ...3 CALibration Subsystem Remote Calibration Procedures 78 U3606A Programmer s Reference ...
Page 132: ...4 CONFigure Subsystem CONFigure DCYCle 118 U3606A Programmer s Reference ...
Page 138: ...5 INITiate Subsystem INITiate CONTinuous 124 U3606A Programmer s Reference ...
Page 174: ...7 MEMory Subsystem MEMory STATe RECall AUTO 160 U3606A Programmer s Reference ...
Page 295: ...SOURce Subsystem 10 SOURce PROTection STATe U3606A Programmer s Reference 281 RST on page 328 ...
Page 296: ...10 SOURce Subsystem SOURce PROTection STATe 282 U3606A Programmer s Reference ...
Page 332: ...14 Root Commands READ 318 U3606A Programmer s Reference ...
Page 350: ...15 IEEE 488 2 Common Commands WAI 336 U3606A Programmer s Reference ...
Page 370: ...Command Quick Reference 356 U3606A Programmer s Reference ...