CALibration Subsystem
3
Remote Calibration Procedures
U3606A Programmer’s Reference
65
Resistance gain adjustment procedures
Follow the steps outlined below. Review the “Test Considerations”
described in the
U3606A User’s and Service Guide
and the
adjustment considerations” on page 59
before beginning this test.
Calibration step
Remote command
Resistance gain adjustment — Short
1
Select the 2-wire resistance measurement.
Connect a shorting plug between the
Ω
(red) and
LO
(black) input terminals.
&
CONF:RES
2
Calibrate the zero point for 2-wire resistance
measurements.
&
CAL:VAL 0
3
Start the calibration.
&
CAL?
Resistance gain adjustment — Open
4
Remove the shorting plug from the input
terminals (all terminals open). Select the
100 M
Ω
range.
&
CONF:RES 100M
5
Calibrate the open point for 2-wire
resistance measurements.
&
CAL:VAL 9.9E+37
6
Start the calibration.
&
CAL?
Resistance gain adjustment — 10 M
Ω
7
Select the 10 M
Ω
range. Input 10 M
Ω
resistance to the
Ω
(red) and
LO
(black)
input terminals.
&
CONF:RES 10M
8
Calibrate the 10 M
Ω
point for 2-wire
resistance measurements.
&
CAL:VAL 10M
9
Start the calibration.
&
CAL?
N O T E
If the zero offset adjustment procedure has been recently
performed prior to the resistance gain calibration procedure, the
adjustment item “Short” and “Open” can be omitted.
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply Programmer s Reference ...
Page 2: ......
Page 3: ......
Page 14: ...XII U3606A Programmer s Reference ...
Page 32: ...1 Introduction to SCPI SCPI Status System 18 U3606A Programmer s Reference ...
Page 58: ...2 CALCulate Subsystem CALCulate NULL OFFSet 44 U3606A Programmer s Reference ...
Page 92: ...3 CALibration Subsystem Remote Calibration Procedures 78 U3606A Programmer s Reference ...
Page 132: ...4 CONFigure Subsystem CONFigure DCYCle 118 U3606A Programmer s Reference ...
Page 138: ...5 INITiate Subsystem INITiate CONTinuous 124 U3606A Programmer s Reference ...
Page 174: ...7 MEMory Subsystem MEMory STATe RECall AUTO 160 U3606A Programmer s Reference ...
Page 295: ...SOURce Subsystem 10 SOURce PROTection STATe U3606A Programmer s Reference 281 RST on page 328 ...
Page 296: ...10 SOURce Subsystem SOURce PROTection STATe 282 U3606A Programmer s Reference ...
Page 332: ...14 Root Commands READ 318 U3606A Programmer s Reference ...
Page 350: ...15 IEEE 488 2 Common Commands WAI 336 U3606A Programmer s Reference ...
Page 370: ...Command Quick Reference 356 U3606A Programmer s Reference ...