CALCulate Subsystem
2
CALCulate:FUNCtion
U3606A Programmer’s Reference
21
CALCulate:FUNCtion?
This query returns a string value that represents the currently selected
calculation function:
AVER
,
DB
,
DBM
,
HOLD
,
LIM
, or
NULL
Parameter
Remarks
•
The CALCulate subsystem must be enabled using the
CALCulate:STATe
command.
•
All calculation functions are not allowed for diode and continuity tests.
•
All calculation functions can be combined with each other with the
exception of the following restrictions:
•
If
DB
and
DBM
are selected, all previously set calculation functions
will be disabled.
•
If
LIMit
is selected, previously set
AVERage
and
HOLD
functions will
be disabled.
•
If
AVERage
is selected, previously set
LIMit
and
HOLD
functions will
be disabled.
•
If
HOLD
is selected, previously set
LIMit
and
AVERage
functions will
be disabled.
•
If
NULL
is selected, previously set
DBM
and
DB
functions will be
disabled.
•
The instrument clears the calculation function selection, reverting to the
default after a Factory Reset (
*RST
command) or an Instrument Preset
(
SYSTem:PRESet
command).
Item
Type
Description
Default value
function
Discrete
AVERage|DB|DBM|HOLD|LIMit|NULL
NULL
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply Programmer s Reference ...
Page 2: ......
Page 3: ......
Page 14: ...XII U3606A Programmer s Reference ...
Page 32: ...1 Introduction to SCPI SCPI Status System 18 U3606A Programmer s Reference ...
Page 58: ...2 CALCulate Subsystem CALCulate NULL OFFSet 44 U3606A Programmer s Reference ...
Page 92: ...3 CALibration Subsystem Remote Calibration Procedures 78 U3606A Programmer s Reference ...
Page 132: ...4 CONFigure Subsystem CONFigure DCYCle 118 U3606A Programmer s Reference ...
Page 138: ...5 INITiate Subsystem INITiate CONTinuous 124 U3606A Programmer s Reference ...
Page 174: ...7 MEMory Subsystem MEMory STATe RECall AUTO 160 U3606A Programmer s Reference ...
Page 295: ...SOURce Subsystem 10 SOURce PROTection STATe U3606A Programmer s Reference 281 RST on page 328 ...
Page 296: ...10 SOURce Subsystem SOURce PROTection STATe 282 U3606A Programmer s Reference ...
Page 332: ...14 Root Commands READ 318 U3606A Programmer s Reference ...
Page 350: ...15 IEEE 488 2 Common Commands WAI 336 U3606A Programmer s Reference ...
Page 370: ...Command Quick Reference 356 U3606A Programmer s Reference ...