SENSe Subsystem
9
[SENSe:]CONTinuity:RANGe[:UPPer]
U3606A Programmer’s Reference
209
[SENSe:]CONTinuity:RANGe[:UPPer]
Syntax
[SENSe:]CONTinuity:RANGe[:UPPer] {<range>|MIN|MAX|DEF}
This command selects the measurement range for continuity tests.
Continuity is a special type of fixed- range 2- wire resistance measurement.
[SENSe:]CONTinuity:RANGe[:UPPer]? [{MIN|MAX}]
This query returns a numeric value that represents the continuity test
range.
Parameter
Remarks
•
Selecting a discrete range will disable autoranging.
•
If the input signal is greater than can be measured on the selected
range (manual ranging), the instrument gives an overload indication.
•
Positive overload: “
+9.37
” or “OL” from the front panel.
•
Negative overload: “
-9.37
” or “–OL” from the front panel.
•
If a range change is in progress (due to the autoranging setting) when
you query the instrument, the value “
+9.37
” will be
returned. Wait for the instrument to select an appropriate range before
querying the instrument again.
•
The instrument is set to 1 k
Ω
, with autoranging disabled, after a
Factory Reset (
*RST
command).
Item
Type
Range of values
Default value
range
Numeric
•
100
Ω
|MIN
•
1 k
Ω
|DEF
•
10 k
Ω
•
100 k
Ω
•
1 M
Ω
•
10 M
Ω
•
100 M
Ω
|MAX
1 k
Ω
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply Programmer s Reference ...
Page 2: ......
Page 3: ......
Page 14: ...XII U3606A Programmer s Reference ...
Page 32: ...1 Introduction to SCPI SCPI Status System 18 U3606A Programmer s Reference ...
Page 58: ...2 CALCulate Subsystem CALCulate NULL OFFSet 44 U3606A Programmer s Reference ...
Page 92: ...3 CALibration Subsystem Remote Calibration Procedures 78 U3606A Programmer s Reference ...
Page 132: ...4 CONFigure Subsystem CONFigure DCYCle 118 U3606A Programmer s Reference ...
Page 138: ...5 INITiate Subsystem INITiate CONTinuous 124 U3606A Programmer s Reference ...
Page 174: ...7 MEMory Subsystem MEMory STATe RECall AUTO 160 U3606A Programmer s Reference ...
Page 295: ...SOURce Subsystem 10 SOURce PROTection STATe U3606A Programmer s Reference 281 RST on page 328 ...
Page 296: ...10 SOURce Subsystem SOURce PROTection STATe 282 U3606A Programmer s Reference ...
Page 332: ...14 Root Commands READ 318 U3606A Programmer s Reference ...
Page 350: ...15 IEEE 488 2 Common Commands WAI 336 U3606A Programmer s Reference ...
Page 370: ...Command Quick Reference 356 U3606A Programmer s Reference ...