4
CONFigure Subsystem
CONFigure[:VOLTage]:AC
84
U3606A Programmer’s Reference
CONFigure[:VOLTage]:AC
Syntax
CONFigure[:VOLTage]:AC
[<range>|AUTO|MAX|MIN|DEF[,{<resolution>|MAX|MIN|DEF}]]
This command first resets all AC voltage measurement parameters and
trigger parameters to their default values. Then, it configures the
instrument for AC voltage measurements but does not initiate the
measurement.
C A U T I O N
•
The maximum range parameter is 750 V
rms
, which is set by MAX.
The rms voltage is waveform dependent. A sine wave is limited to
750 V
ac
(rms), but a 1000 V
pk
square wave is safe. Connections to
AC MAINS are further limited to CAT II (300V).
•
See the “Safety Information” section in the
U3606A User's and
Service Guide
for a complete discussion of the safety features, and
the procedures for safe operation of this instrument.
N O T E
The
CONFigure[:VOLTage]:AC
command does not place the
instrument in the “wait-for-trigger” state. Use the
INITiate[:IMMediate]
or
READ?
command in conjunction with the
CONFigure[:VOLTage]:AC
to place the instrument in the
“wait-for-trigger” state.
Summary of Contents for U3606A
Page 1: ...Keysight U3606A Multimeter DC Power Supply Programmer s Reference ...
Page 2: ......
Page 3: ......
Page 14: ...XII U3606A Programmer s Reference ...
Page 32: ...1 Introduction to SCPI SCPI Status System 18 U3606A Programmer s Reference ...
Page 58: ...2 CALCulate Subsystem CALCulate NULL OFFSet 44 U3606A Programmer s Reference ...
Page 92: ...3 CALibration Subsystem Remote Calibration Procedures 78 U3606A Programmer s Reference ...
Page 132: ...4 CONFigure Subsystem CONFigure DCYCle 118 U3606A Programmer s Reference ...
Page 138: ...5 INITiate Subsystem INITiate CONTinuous 124 U3606A Programmer s Reference ...
Page 174: ...7 MEMory Subsystem MEMory STATe RECall AUTO 160 U3606A Programmer s Reference ...
Page 295: ...SOURce Subsystem 10 SOURce PROTection STATe U3606A Programmer s Reference 281 RST on page 328 ...
Page 296: ...10 SOURce Subsystem SOURce PROTection STATe 282 U3606A Programmer s Reference ...
Page 332: ...14 Root Commands READ 318 U3606A Programmer s Reference ...
Page 350: ...15 IEEE 488 2 Common Commands WAI 336 U3606A Programmer s Reference ...
Page 370: ...Command Quick Reference 356 U3606A Programmer s Reference ...