Front-Panel Operation
3
Keysight E3633A and E3634A User’s Guide
97
Calibration count
You can determine the number of times that your power supply has been
calibrated. Your power supply was calibrated before it left the factory. When you
receive your power supply, read the count to determine its initial value.
The calibration count feature can be performed from the remote interface only.
– The calibration count is stored in
non-volatile
memory, and does not
change when power has been off or after a remote interface reset.
– The calibration count increments up to a maximum of 32,767 after which it
wraps around to 0. Since the value increments by one for each calibration
point, a complete calibration will increase the value by 5 counts.
–
Remote interface operation:
CAL:COUN?
Query the number of times of calibration