Front-Panel Operation
3
Keysight E3633A and E3634A User’s Guide
63
Programming Overcurrent Protection
Overcurrent protection guards the load against output currents that reach a
specified value greater than the programmed protection level. It is accomplished
by programming the output current to zero.
The following steps show how to set the overcurrent protection trip level, how to
check OCP operation and how to clear overcurrent condition.
– Front-panel operation:
Setting the OCP level and enable the OCP circuit
1 Turn on the power supply.
The power supply will go into the
power-on / reset
state; the output is disabled
(the
OFF
annunciator turns on); the 8V/20A
[1]
or 25V/7A
[2]
range is selected
(the
8V
or
25V
annunciator turns on); and the knob is selected for
voltage
control.
2 Enable the output.
The
OFF
annunciator turns off and the display will go to the meter mode.
3 Enter the OCP menu and set the trip level.
You will see the above message on the display when you enter the OCP menu.
Adjust the knob for the desired OCP trip level.
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
LEVEL 2
2
.0 A (E3633A)
LEVEL
7
.5 A (E3634A)
Summary of Contents for E3633A
Page 1: ...Keysight E3633A and E3634A DC Power Supplies User s Guide ...
Page 10: ...10 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...20 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 133: ...Remote Interface Reference 4 Keysight E3633A and E3634A User s Guide 133 SCPI Status System ...