1
General Information
34
Keysight E3633A and E3634A User’s Guide
Front-panel voltage and current limit settings
You can set the voltage and current limit values from the front panel using the
following method.
1
Select the desired range using the range selection keys after turning on the
power supply.
2
Press the
key to show the limit values on the display.
3
Move the blinking digit to the appropriate position using the resolution
selection keys and change the blinking digit value to the desired voltage limit
by turning the control knob. If the display limit times out, press the
key
again.
4
Set the knob to current control mode by pressing the
key.
5
Move the blinking digit to the appropriate position using the resolution
selection keys and change the blinking digit value to the desired current limit
by turning the control knob.
6
Press the
key to enable the output. After about 5 seconds, the display will
go to output monitoring mode automatically to display the voltage and current
at the output or the display will go to output monitoring mode immediately by
pressing the
key again.
Use the voltage/current adjust selection key, the resolution selection keys, and the control knob to
change the voltage and current limit values.
NOTE
All front panel keys and controls can be disabled with remote interface
commands. The Keysight E3633A and Keysight E3634A must be in "Local" mode
for the front panel keys and controls to function.
Summary of Contents for E3633A
Page 1: ...Keysight E3633A and E3634A DC Power Supplies User s Guide ...
Page 10: ...10 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...20 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 133: ...Remote Interface Reference 4 Keysight E3633A and E3634A User s Guide 133 SCPI Status System ...