5
Error Messages
168
Keysight E3633A and E3634A User’s Guide
712
Bad DAC cal data
The specified DAC calibration values (
CAL:VOLT
or
CAL:CURR
) are out of range.
Note that the new calibration constants are not stored in the non-volatile
memory.
713
Bad readback cal data
The specified readback calibration values (
CAL:VOLT
or
CAL:CURR
) are out of
range. Note that the new calibration constants are not stored in the non-volatile
memory.
714
Bad OVP cal data
The overvoltage protection calibration constant is out of range. Note that the
new calibration constants are not stored in the non-volatile memory.
715
Bad OCP cal data
The overcurrent protection calibration constant is out of range. Note that the
new calibration constants are not stored in the non-volatile memory.
716
Bad DAC DNL error correction data
Invalid data measured during the calibration for DAC differential nonlinearity
error correction.
717
Cal OVP or OCP status enabled
Overvoltage protection status or overcurrent protection status is enabled. You
must set both overvoltage and overcurrent protection status to OFF before and
during the calibration.
718
Gain out of range for Gain Error Correction
The slope of the DAC gain is out of range. Hardware fails.
740
Cal checksum failed, secure state
Summary of Contents for E3633A
Page 1: ...Keysight E3633A and E3634A DC Power Supplies User s Guide ...
Page 10: ...10 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...20 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 133: ...Remote Interface Reference 4 Keysight E3633A and E3634A User s Guide 133 SCPI Status System ...