CALibration:VOLTage[:DATA] <numeric value>
This command can only be used after calibration is unsecured and the output
state is ON. It enters a voltage value that you obtained by reading an external
meter. You must first select the minimum calibration level (
CAL:VOLT:LEV MIN
)
for the value being entered. You must then select the middle and maximum
calibration levels (
CAL:VOLT:LEV MID
and
CAL:VOLT:LEV MAX
) for the value
being entered. Three successive values must be selected and entered. The
power supply then computes new voltage calibration constants. These
constants are then stored in
non-volatile
memory.
CALibration:VOLTage:LEVel {MINimum|MIDdle|MAXimum}
This command can only be used after calibration is unsecured and the output
state is ON. It sets the power supply to a calibration point that is entered with
CAL:VOLT
command. During calibration, three points must be entered and the
low-end point (MIN) must be selected and entered first.
CALibration:VOLTage:PROTection
This command calibrates the overvoltage protection circuit of the power
supply. It takes about 10 seconds to execute the command. The calibration
must be unsecured and the output be opened before calibrating the
overvoltage protection circuit. The power supply automatically performs the
calibration and stores the new overvoltage constant in
nonvolatile
memory.
Notice that voltage calibration precedes before sending this command.
Summary of Contents for E3633A
Page 1: ...Keysight E3633A and E3634A DC Power Supplies User s Guide ...
Page 10: ...10 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...20 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 133: ...Remote Interface Reference 4 Keysight E3633A and E3634A User s Guide 133 SCPI Status System ...