8
Specifications
198
Keysight E3633A and E3634A User’s Guide
Performance Specifications
Table 8-1
Performance specifications
Parameter
Keysight E3633A
Keysight E3634A
Output Ratings
(@ 0
°C- 40
°C)
Low Range
0 to +8 V/ 0 to 20 A 0 to 25 V/ 0 to 7 A
High Range
0 to +20V/ 0 to 10 A 0 to+50 V/ 0 to 4 A
Programming Accuracy
[a]
12 months (@ 25 °C ± 5 °C),
±(% of offset)
[a] Accuracy specifications are after an 1-hour warm-up with no load and calibration at 25 °C.
Voltage
0.05% + 10 mV
Current
0.2% + 10 mA
Readback Accuracy
[a][b]
12 months (over GPIB and
RS-232 or front panel with
respect to actual output @ 25 °C
± 5 °C), ±(% of offset)
[b] This specification may degrade when the unit is subjected to an RF field >= 3V/meter.
Voltage
0.05% + 5 mV
Current
0.15% + 5 mA
Ripple and Noise
(with outputs ungrounded, or
with either output terminal
grounded, 20 Hz to 20 MHz)
Normal mode voltage
<0.35 mV rms and
3 mV p-p
<0.5 mV rms and
3 mV p-p
Normal mode current
<2 mA rms
Common mode current <1.5 uA rms
Load Regulation
,
±(% of offset)
Voltage
<0.01% + 2 mV
Current
<0.01% + 250
μ
A
Line Regulation
,
±(% of offset)
Voltage
<0.01% + 2 mV
Current
<0.01% + 250
μ
A
Programming Resolution
Voltage
1 mV
3 mV
Current
1 mA
0.5 mA
Readback Resolution
Voltage
0.5mV
1.5 mV
Current
1 mA
0.5 mA
Front Panel Resolution
Voltage
1 mV
Current
1 mA (< 10A), 10mA (. 10A)
Summary of Contents for E3633A
Page 1: ...Keysight E3633A and E3634A DC Power Supplies User s Guide ...
Page 10: ...10 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...20 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 133: ...Remote Interface Reference 4 Keysight E3633A and E3634A User s Guide 133 SCPI Status System ...