4
Remote Interface Reference
136
Keysight E3633A and E3634A User’s Guide
The standard event register is cleared when:
– You execute the
*CLS
(clear status) command.
– You query the event register using the
*ESR?
(Event Status register) command.
For example, 28 (4 + 8 + 16) is returned when you have queried the status of the
Standard Event register, QYE, DDE, and EXE conditions have occurred.
The standard event enable register is cleared when:
– You execute the
*ESE 0
command.
– You turn on the power and have previously configured the power supply using
the
*PSC 1
command.
– The enable register will not be cleared at power-on if you have previously
configured the power supply using the
*PSC 0
command.
The status byte register
The Status Byte summary register reports conditions from the other status
registers. Query data that is waiting in the power supply’s output buffer is
immediately reported through the “Message Available” bit (bit 4) of Status Byte
register. Bits in the summary register are not latched. Clearing an event register
will clear the corresponding bits in the Status Byte summary register. Reading all
messages in the output buffer, including any pending queries, will clear the
message available bit.
Table 4-5
Bit-Definitions - Status Byte Summary Register
Bit
Decimal Value Definition
0-2 Not Used 0
Always set to 0.
3
QUES
8
One or more bits are set in the questionable status register (bits must
be “enabled” in the enable register).
4
MAV
16
Data is available in the power supply output buffer.
5
ESB
32
One or more bits are set in the standard event register (bits must be
“enabled” in the enable register).
6
RQS
64
The power supply is requesting service (serial poll).
7
Not Used 0
Always set to 0.
Summary of Contents for E3633A
Page 1: ...Keysight E3633A and E3634A DC Power Supplies User s Guide ...
Page 10: ...10 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...20 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 133: ...Remote Interface Reference 4 Keysight E3633A and E3634A User s Guide 133 SCPI Status System ...