Front-Panel Operation
3
Keysight E3633A and E3634A User’s Guide
75
Error conditions
When the front-panel ERROR annunciator turns on, one or more command syntax
or hardware errors have been detected. A record of up to 20 errors can be stored
in the power supply’s error queue. See
complete listing of the errors.
– Errors are retrieved in first-in-first-out (FIFO) order. The first error returned
is the first error that was stored. Errors are cleared as you read them over
the remote interface. When you have read all errors from the queue, the
ERROR
annunciator turns off. The power supply beeps once each time an
error is generated.
– If more than 20 errors have occurred when you operate the power supply
over the remote interface, the last error stored in the queue (the most
recent error) is replaced with -350,
“Queue overflow”
. No additional errors
are stored until you remove errors from the queue. If no errors have
occurred when you read the error queue, the power supply responds with
+0,
“No error”
over the remote interface or “NO ERRORS” from the front
panel.
– The error queue is cleared when power has been off or after a
*CLS
(clear
status) command has been executed. The
*RST
(reset) command does not
clear the error queue.
Summary of Contents for E3633A
Page 1: ...Keysight E3633A and E3634A DC Power Supplies User s Guide ...
Page 10: ...10 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...20 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 133: ...Remote Interface Reference 4 Keysight E3633A and E3634A User s Guide 133 SCPI Status System ...