Error Messages
5
Keysight E3633A and E3634A User’s Guide
161
-158
String data not allowed
A character string was received but is not allowed for the command. Check the
list of parameters to verify that you have used a valid parameter type.
Example:
TRIG:DEL ‘zero’
-160 to -168
Block data errors
The power supply does not accept block data.
-170 to -178
Expression errors
The power supply does not accept mathematical expressions.
-211
Trigger ignored
A Group Execute Trigger (GET) or
*TRG
was received but the trigger was
ignored. Make sure that the trigger source should be selected to the bus and the
trigger subsystem should be initiated by
INIT[:IMM]
command.
-213
Init ignored
An
INITiate
command was received but could not be executed because a
measurement was already in progress. Send a device clear to halt a
measurement in progress and place the power supply in the “idle” state.
-221
Settings conflict
Indicates that a legal program data element was parsed but could not be
executed due to the current device state.
-222
Data out of range
A numeric parameter value is outside the valid range for the command.
Example:
TRIG:DEL -3
Summary of Contents for E3633A
Page 1: ...Keysight E3633A and E3634A DC Power Supplies User s Guide ...
Page 10: ...10 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 20: ...20 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 133: ...Remote Interface Reference 4 Keysight E3633A and E3634A User s Guide 133 SCPI Status System ...