Keysight M8070A Programming Guide
431
SCPI Command Reference
5
:PLUGin:PSWEep:RUN:MESSage?
Syntax
:PLUGin:PSWEep:RUN:MESSage? 'Identifier'
Input
Parameters
‘Identifier’: Specify the plugin name. For e.g. ‘ParameterSweep1’
Description
This command returns a string describing the state of the parameter
sweep addressed by the measurement name identifier. Possible states
include; Not Started, Running, Suspended, Finished, Error, or Stopped.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A, M8045A
and M8046A.
Example
:PLUG:PSWE:RUN:MESS? 'ParameterSweep1'
:PLUGin:PSWEep:RUN:PROGress?
Syntax
:PLUGin:PSWEep:RUN:PROGress? 'Identifier'
Input
Parameters
‘Identifier’: Specify the plugin name. For e.g. ‘ParameterSweep1’
Return Range
0 to 1
Description
This command returns a number in the range of 0.0 to 1.0 to indicate the
progress of an parameter sweep measurement addressed by the
measurement name identifier. 0.0 indicates that the measurement has not
started and 1.0 indicates the measurement is finished.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A, M8045A
and M8046A.
Example
:PLUG:PSWE:RUN:PROG? 'ParameterSweep1'
:PLUGin:PSWEep:RUN:STATus?
Syntax
:PLUGin:PSWEep:RUN:STATus? 'Identifier'
Input
Parameters
‘Identifier’: Specify the plugin name. For e.g. ‘ParameterSweep1’
Return Range
0|1
Description
This command returns the running status of a parameter sweep
measurement addressed by the measurement name identifier. 0 indicates