:PLUGin:PSWEep:ACQuisition:SRTime[?]
Syntax
:PLUGin:PSWEep:ACQuisition:SRTime 'Identifier', <NRf>
:PLUGin:PSWEep:ACQuisition:SRTime? 'Identifier'
Input
Parameters
‘Identifier’: Specify the plugin name. For e.g. ‘ParameterSweep1’
Description
This command sets the relaxing time between the sweep before reading
the BER. The relaxing time can be between 0 sec to 60 sec.
This query returns the current set relaxing time.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A, M8045A
and M8046A.
Example
:PLUG:PSWE:ACQ:SRT 'ParameterSweep1', 1E-2
:PLUGin:PSWEep:ACQuisition:SFRom[?]
Syntax
:PLUGin:PSWEep:ACQuisition:SFRom 'Identifier', <NRf>
:PLUGin:PSWEep:ACQuisition:SFRom? 'Identifier',
Input
Parameters
‘Identifier’: Specify the plugin name. For e.g. ‘ParameterSweep1’'
Description
This command sets the value from where the sweep needs to be
performed for that particular parameter.
This query returns the current sweep value.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A, M8045A
and M8046A.
Example
:PLUG:PSWE:ACQ:SFR 'ParameterSweep1', 1
:PLUGin:PSWEep:ACQuisition:STO[?]
Syntax
:PLUGin:PSWEep:ACQuisition:STO 'Identifier', <NRf>
:PLUGin:PSWEep:ACQuisition:STO? 'Identifier'
Input
Parameters
‘Identifier’: Specify the plugin name. For e.g. ‘ParameterSweep1’