5
[:SOURce]:INTerference:ISYMbol:ILOSs{1:2}[?]
Syntax
[:SOURce]:INTerference:ISYMbol:ILOSs1 'Identifier', <NRf>
[:SOURce]:INTerference:ISYMbol:ILOSs2 'Identifier', <NRf>
[:SOURce]:INTerference:ISYMbol:ILOSs1? 'Identifier'
[:SOURce]:INTerference:ISYMbol:ILOSs2? 'Identifier'
Input
Parameters
Identifier: 'M*.DataOut', 'M*.DataOut1', or 'M*.DataOut2'
Description
This command sets the insertion loss at one or two discrete user definable
frequencies, f1 and f2 of the inter symbol interference (ISI) trace.
This SCPI is applicable for M8041A, M8051A, or M8062A.
Range
0.0dB to -25.0dB.
Example
:SOUR:INT:ISYM:ILOS1 'M2.DataOut', -9.8
[:SOURce]:INTerference:ISYMbol:PRESet[?]
Syntax
[:SOURce]:INTerference:ISYMbol:PRESet 'Identifier', <CUSTom | P1 | P2 |
P3 | P4 | P5 | P6 | P7 | P8 | P9 | P10 | P11 | P12 | P13 | P14 | P15 | P16 | P17>
[:SOURce]:INTerference:ISYMbol:PRESet? 'Identifier'
Input
Parameters
‘identifier’: 'M*.DataOut1', 'M*.DataOut2', or 'M*.DataOut'
Description
Selects an inter symbol interference (ISI) preset for the specific
application.
This command should ideally not be sent in one PM (SCPI - Programm
Message) with following commands…:MODE; …:FREQuency; …:ILOSs.
This SCPI is applicable for M8041A, M8051A and M8062A.
For M8062A the presets are available up to P11 only; please refer to the