376
Keysight M8070A Programming Guide
5
SCPI Command Reference
:PLUGin:JTOLerance:MSETup:FREQuency:STOp[?]
Syntax
:PLUGin:JTOLerance:MSETup:FREQuency:STOp ‘identifier’, <NRf>
:PLUGin:JTOLerance:MSETup:FREQuency:STOp? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<NRf>: Set the stop frequency.
Return Range
100 Hz to 500 MHz
Description
This command is used to set the stop (maximum) modulation frequency for
the measurement specified by the measurement name identifier. The
default is 100 MHz.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example
:PLUG:JTOL:MSET:FREQ:STO ‘MyMeasurement’, 100MHz
:PLUGin:JTOLerance:MSETup:MTPoints[?]
Syntax
:PLUGin:JTOLerance:MSETup:MTPoints[?] ‘identifier’, <NRf>
Input
Parameters
‘identifier’: Specify the measurement name.
Return Range
0 | 1 OFF | ON
Description
This command selects either template points or specified range.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example
:PLUGin:JTOLerance:MSETup:MTPoints ‘MyMeasurement’
:PLUGin:JTOLerance:MSETup:NPOints[?]
Syntax
:PLUGin:JTOLerance:MSETup:NPOints ‘identifier’, <NRf>
:PLUGin:JTOLerance:MSETup:NPOints? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<NRf>: Set the number of measurement points.
Return Range
1 to 100