332
Keysight M8070A Programming Guide
5
SCPI Command Reference
:PLUGin:OTIMing:FETCh[:JITTer][:TOTal]:UNCertainty?
Syntax
:PLUGin:OTIMing:FETCh[:JITTer][:TOTal]:UNCertainty?
‘identifier’[,‘Location’]
Input
Parameters
‘identifier’: Specify the measurement name.
[‘Location’]: Specify location (optional).
Return Range
<NR3>
Description
This command returns the total jitter uncertainty associated with a specific
measurement name identifier. The returned value is only valid if the
“Sample Delay Optimization” is set to “Fast Total Jitter” (FTJ) using the
:PLUG:OTIM:ACQ:OPT command.
This SCPI is applicable for M8041A, M8051A, M8062A and M8046A.
A measurement run on a location identifier (for example, ‘M1.DataIn1’) will
return following result.
Example
:PLUG:OTIM:FETC:UNC? ‘MyOutputTiming’,‘M1.DataIn1’
:PLUGin:OTIMing:FETCh:DATA?
Syntax
:PLUGin:OTIMing:FETCh:DATA? ‘identifier’[,‘Location’]
Input
Parameters
‘identifier’: Specify the measurement name.
[‘Location’]: Specify location (optional).
Description
This command returns the raw data of the output timing measurement
associated with a specific measurement name identifier.
A measurement run on a location identifier (for example, ‘M1.DataIn1’) will
return following results.
This SCPI is applicable for M8041A, M8051A and and M8062A.
Example
:PLUG:OTIM:FETC:DATA? ‘Output Timing 1’,‘M1.DataIn1’
"M1.DataIn1",1420788572.96819,499999011,500000989,249999011,250
000000,-0.6035,"UNKNOWN", ...
The first item in the comma separated list is the name of the location for
which the data is being queried. Rest of the values in the comma
separated list for each measurement point are Time Stamp, Compared