300
Keysight M8070A Programming Guide
5
SCPI Command Reference
:PLUGin:ERATio:STOP
Syntax
:PLUGin:ERATio:STOP ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name to stop.
Description
This command stops an error ratio measurement addressed by the
measurement name identifier.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example
The following example stops an error ratio measurement with the
measurement name identifier called ‘MyMeasurement’:
:PLUG:ERAT:STOP ‘MyMeasurement’
:PLUGin:ERATio:SHOW:SGLegends
Syntax
:PLUGin:ERATio:SHOW:SGLegends 'Identifier', <0|1|OFF|ON>
:PLUGin:ERATio:SHOW:SGLegends? 'Identifier'
Input
Parameters
Identifier: 'MyMeasurement'
<0|1|OFF|ON>: Specify the show/hide state for graph legends.
Return Range
0|1
Description
This command shows/hides the graph legends.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example
:PLUGin:ERATio:SHOW:SGLegends 'MyMeasurement', 1
:PLUGin:ERATio:GRAPh?
Syntax
:PLUGin:ERATio:GRAPh? ‘identifier’
[,DesiredWidth][,DesiredHeight][,<CURR | WHIT>][,<PNG>][,1 |
0][,’TabName’]
Input
Parameters
‘identifier’: Specify measurement name.
Other optional parameters are:
[,DesiredWidth]: Specify the desired width of the image.
[,DesiredHeight]: Specify the desired height of the image.