250
Keysight M8070A Programming Guide
5
SCPI Command Reference
:INPut:CDR:FIRSt:TDENsity[?]
Syntax
:INPut:CDR:FIRSt:TDENsity ‘identifier’, <NRf>
:INPut:CDR:FIRSt:TDENsity? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn1' or 'M*.DataIn2'
<NRf>: Set first order transition density.
Return Range
25% to 100%
Description
This command is used to set the first order transition density of the
incoming signal from 25% to 100%. The default is 50%.
This SCPI is applicable for M8041A and M8051A.
Example
:INP:CDR:FIRS:TDEN ‘M1.DataIn1’, 25
:INPut:CDR:FIRSt:LBANdwidth[?]
Syntax
:INPut:CDR:FIRSt:LBANdwidth ‘identifier’, <NRf>
:INPut:CDR:FIRSt:LBANdwidth? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn1' or 'M*.DataIn2'
<NRf>: Set first order loop bandwidth of the CDR.
Return Range
102 kHz to 20 MHz
Description
This command is used to set the first order loop bandwidth of the CDR.
The default is 2 MHz. Acceptable units are Hz, kHz, MHz and exponents
(for example, 4E6 is the same as 4 MHz).
This SCPI is applicable for M8041A and M8051A.
Example
:INP:CDR:FIRS:LBAN ‘M1.DataIn1’, 4MHz