SCPI Command Reference
:PLUGin:JTOLerance:INSTruments:GENerator[?]
Syntax
:PLUGin:JTOLerance:INSTruments:GENerator ‘identifier’,<location-string>
:PLUGin:JTOLerance:INSTruments:GENerator? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<location-string>: Specify the location identifier or group name identifier.
Return Range
<location or group name>
Description
This command is used to select the pattern generator module/channel for
the jitter tolerance measurement. The location-string refers to either a
location name or a group name identifier associated with a measurement
name identifier.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example
:PLUG:JTOL:INST:GEN ‘MyMeasurement’,‘M1.DataOut1’
:PLUGin:JTOLerance:INSTruments:ANALyzer[?]
Syntax
:PLUGin:JTOLerance:INSTruments:ANALyzer ‘identifier’,<location-string>
:PLUGin:JTOLerance:INSTruments:ANALyzer? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<location-string>: Specify the location identifier or group name identifier.
Return Range
<location or group name>
Description
This command is used to select the analyzer (error detector)
module/channel for the jitter tolerance measurement. The location-string
refers to either a location name or a group name identifier associated with
a measurement name identifier.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example