190
Keysight M8070A Programming Guide
5
SCPI Command Reference
[:SOURce]:INTerference[:LEVel]:CMODe[:STATe][?]
Syntax
[:SOURce]:INTerference[:LEVel]:CMODe[:STATe] ‘identifier’, <0|1|ON|OFF>
[:SOURce]:INTerference[:LEVel]:CMODe[:STATe]? ‘identifier’
Input
Parameters
‘identifier’: ‘M*.DATAOUT1’, ‘M*.DataOut2’, or ‘M*.DataOut’
<0|1|ON|OFF>: Turn common mode interference on/off.
Return Range
0|1
Description
This command turns common mode interference on/off.
This SCPI is applicable for M8041A, M8051A, M8061A and M8062A.
Example
:INT:CMOD ‘M2.DataOut1’, on
[:SOURce]:INTerference[:LEVel]:CMODe:AMPLitude[?]
Syntax
[:SOURce]:INTerference[:LEVel]:CMODe:AMPLitude ‘identifier’, <NRf>
[:SOURce]:INTerference[:LEVel]:CMODe:AMPLitude? ‘identifier’
Input
Parameters
‘identifier’: ‘M*.DATAOUT1’ or ‘M*.DataOut2’
<NRf>: Set common mode interference amplitude.
Return Range
0 mV to 320 mV
Description
This command sets the amplitude of the common mode interference in
volts. Acceptable units include mV, V and exponents (for example, 100E-3
is the same as 100 mV).
This SCPI is applicable for M8041A and M8051A.
Example