B-18
Calibration Reference
:DATE?
(:CALibration:PROTected:CARD1:DATE?)
(:CALibration:PROTected:CARD2:DATE?)
Purpose
To request the Model 7700 calibration date.
Format
:cal:prot:card1:date?
:cal:prot:card2:date?
Response
<year>, <month>, <day>
Description
The
:CARD1:DATE?
and
:CARD2:DATE?
queries allow you to read
back the calibration date from a Model 7700 in Slot 1 and Slot 2
respectively.
NOTE
The card calibration date is automatically set to the Model 2700 real time clock date
when the card is calibrated.
Example
:CAL:PROT:CARD1:DATE?
Request card 1 cal date.
:INIT
(:CALibration:PROTected:CARD1:INITiate)
Purpose
To initiate Model 7700 calibration procedures.
Format
:cal:prot:card1:init
Parameter
None
Description
The
:INIT
command enables Model 7700 calibration when performing
these procedures over the bus. This command must be sent to the unit after
sending the
:CODE
command, but before performing Model 7700
calibration.
Example
:CAL:PROT:CARD1:INIT
Initiate 7700 calibration.
Summary of Contents for 2700
Page 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Page 16: ......
Page 17: ...1 PerformanceVeri cation...
Page 48: ...1 32 Performance Verification...
Page 49: ...2 Calibration...
Page 72: ...2 24 Calibration...
Page 73: ...3 RoutineMaintenance...
Page 81: ...4 Troubleshooting...
Page 99: ...5 Disassembly...
Page 106: ...5 8 Disassembly...
Page 107: ...6 ReplaceableParts...
Page 121: ...A Speci cations...
Page 136: ...A 16 Specifications...
Page 137: ...B CalibrationReference...
Page 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Page 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Page 173: ......
Page 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...