![Keithley 2700 Service Manual Download Page 41](http://html1.mh-extra.com/html/keithley/2700/2700_service-manual_3944616041.webp)
Performance Verification
1-25
Verifying resistance
Check resistance by connecting accurate resistance values to the Model 7700 and verifying
that its resistance readings are within the specified limits.
CAUTION
Do not apply more than 300V between plug-in module INPUT or SENSE
H and L terminal, or between any adjacent channels, or instrument dam-
age could occur.
Follow these steps to verify resistance accuracy:
1.
Using shielded Teflon or equivalent cables in a 4-wire configuration, connect the
Model 7700 CH1 H and L INPUT terminals, and CH11 H and L SENSE terminals to
the calibrator as shown in
Figure 1-12
.
Figure 1-12
Connections for Model 7700 resistance verification (100
Ω
to 10M
Ω
ranges)
2.
Install the Model 7700 in Slot 1 of the Model 2700, then turn on the power, and allow
the unit to warm up for two hours before proceeding. Be sure the front panel INPUTS
switch is set to the REAR position.
3.
Set the calibrator for 4-wire resistance with external sense on.
4.
Select the Model 2700 4-wire resistance function by pressing the
Ω
4 key. Close Chan-
nel 1 by pressing the CLOSE key and keying in 101.
5.
Set the Model 2700 for the 100
Ω
range, and make sure the FILTER is on. Enable
OCOMP (offset-compensated ohms) for the 100
Ω
range test. (Press SHIFT then
OCOMP.)
6.
Recalculate reading limits based on actual calibrator resistance values.
H L H L
AMPS
H L H L
H L H L
H L H L
LO
CH21 CH22
CH11 CH12 CH13 CH14 CH15 CH16
H L H L H L H L
CH17 CH18 CH19 CH20
SENSE
(OHMS, 4 WIRE)
INPUT
(V, 2 WIRE)
H L H L H L H L
CH7
CH8 CH9 CH10
H L H L H L H L
H L H L H L H L
INPUT SENSE CH1
CH2
CH3
CH4
CH5
CH6
CH1
CH11
Output
HI
Output
LO
Resistance Calibrator
Note: Use shielded, low-thermal cables
to minimize noise. Enable or disable
calibrator external sense as indicated
in procedure.
Model 7700
Sense HI
Sense LO
Summary of Contents for 2700
Page 1: ...Contains Servicing Information Model 2700 Multimeter DataAcquisition System Service Manual...
Page 16: ......
Page 17: ...1 PerformanceVeri cation...
Page 48: ...1 32 Performance Verification...
Page 49: ...2 Calibration...
Page 72: ...2 24 Calibration...
Page 73: ...3 RoutineMaintenance...
Page 81: ...4 Troubleshooting...
Page 99: ...5 Disassembly...
Page 106: ...5 8 Disassembly...
Page 107: ...6 ReplaceableParts...
Page 121: ...A Speci cations...
Page 136: ...A 16 Specifications...
Page 137: ...B CalibrationReference...
Page 161: ...C CalibrationProgram Appendix C Page 1 Friday February 18 2000 7 45 AM...
Page 172: ...Service fm Page 2 Friday February 18 2000 7 44 AM...
Page 173: ......
Page 174: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...